2.3.9
Storing Results
Press [STORE] key while the test result is displayed.
The display shows:
We advise you to enter a numeric code (up to 6 digits). A code no. can
contain references to the operator or the sampling location.
e.g. [Shift] + [0] [0] [0] [1] [2] [3]
Confirm with [ ] key.
If a code number is not necessary, confirm by pressing [ ] key direction.
(The assignment for the Code No. is 0 automatically).
The display shows:
The test results is then shown again.
The entire data set is stored with date, time, Code No., method and test
result.
Note:
The display also shows the number of free data sets.
If there are fewer than 30 data sets are free, the display shows:
Clear the memory as soon as possible (see Chapter 2.4.3 Delete Stored
Results). If memory capacity is full, it is not possible to save additional test
results.
2.3.10
Printing Results – T2o PC-Link (Optional)
If the PC-Link module is switched on (see Chapter 2.5 Data Transfer) and the printer is connected, it
is possible to print out the test results without saving beforehand.
Press [F3] key.
The entire data set is printed with date, time Code No., method and test
result.
Example:
The test No. is an internal number that is set automatically if a test result is
stored. It appears only on the printed copy.
Saved!
Code-No.:
_ _ _ _ _ _
Storage: 999
free records left
Storage: only
free records left
100 Chlorine T
0.02-
6 mg/l Cl2
Profi-Mode: no
2012-06-15 19:30:22
Test No.: 1
Code-Nr.: 007
4.80 mg/l Cl2
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...