Prepare the first standard and press [TEST] key.
The display shows the input value and measured absorption value.
Confirm with [ ] key.
The display shows:
Enter the concentration of the second standard
e.g. [Shift] + [0] [.][1]
Press [ESC] to delete a digit.
Press [F1] key to reset numerical input.
Confirm with [ ] key.
The display shows:
Prepare the second standard and press [TEST] key.
The display shows the input value and measured absorption value.
Confirm with [ ] key.
The display shows:
Note:
- Perform as described above to measure further standards.
- The minimum number of measured standards is 2.
- The maximum number of measured standards is 14 (S1 to S14).
If all required standards or the maximum value of 14 standards are
measured, press [Store] key.
The display shows:
The instrument goes back to the mode menu automatically.
Now the concentration is stored in the instrument and can be recalled by
entering its method number or selecting it from the displayed method list.
TIP:
Save all your concentration data in a written form because, in case of power outage (e.g.
changing the battery), all concentration data will lost and must be entered again.
<User concentr.>
S1: 0.05 mg/l
mAbs: 12
<User concentr.>
S1 accepted
S2: + _____
│ ESC │F1
<User concentr.>
S2: 0.10 mg/l
prepare Test
press TEST
<User concentr.>
S1: 0.10 mg/l
mAbs: 150
<User concentr.>
S2 accepted
S3: + ______
│ ESC │ F1 │ Store
<User concentr.>
stored!
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...