Instead of zeroing the instrument, press [MODE], [Shift] + [4] [6].
Confirm with [ ] key.
The display shows:
Press [Shift] + [1] keys to delete user calibration.
Press [Shift] + [0] keys to keep the valid user calibration.
The instrument goes back to zero-query automatically.
2.4.5
Lab Function
Reduced Operator Guidance – ‘Profi-Mode’
This function may be used for routine analyses with many samples of one method. The following
information is always stored in the methods:
a) Method
b) Range
c) Date and time
d) Differentiation of results
e) Detailed operator instruction
f) Compliance with reaction periods
If the Profi-Mode is active, the photometer provides only a minimum of operator instructions. The
criteria specified above in d, e and f are no longer included.
Press [MODE], [Shift] + [5] [0] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [0] keys to switch the Profi-Mode off.
Press [Shift] + [q] keys to switch the Profi-Mode on.
The display shows either
<User calibration>
100
Chlorine T
0.02-6 mg/l Cl2
1.00 mg/l free Cl2
Clear user calibration?
YES: 1, NO: 0
<Profi-Mode>
ON: 1, OFF: 0
<Profi-Mode>
Switched off
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...