Prior to Use
Remove the protective cap from the bottom of the sensor.
3.1 Power On/Off
•
To turn the meter on, press the ON/OFF key.
•
To turn the meter off, press and hold the ON/OFF key for 3 seconds,
•
The meter will turn off automatically if there are no key presses within a specified time
period. To disable the auto-off function, go to the Setup Menu section.
3.2 pH Measurement
1. Press the MODE key until the display shows the
indicator.
2. Rinse the pH electrode with distilled water to clean.
3. Immerse the electrode in the sample solution, and stir the solution gently.
4. Wait for the reading to stabilise, and record the value shown on the display.
3.3 ORP Measurement
The HC1000 has two millivolt measurement modes.
ABSOLUTE MILLIVOLT MODE:
•
Press the MODE key until the display shows the measurement unit “mV”; the meter
enters absolute millivolt measurement mode.
RELATIVE MILLIVOLT MODE:
•
Press the MODE key until the display shows the
indicator; the meter enters relative
millivolt measurement mode.
Select one of the above modes. Immerse the electrode in the sample solution, and stir the
solution gently. Wait for the reading to stabilise, and record the value shown on the display.
mV
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...