Color
Value
Red
Extraction or analysis failed
Yellow
Warnings for extraction or analysis
“FAILURE: Analysis Fail Bad Data; Error
Number=nnnnn WARNING…"
Green
Successful extraction (with no analysis intended), or
successful extraction and analysis.
(Sequence analysis only): The Score column represents the quality values for an
entire sequence. Quality values are assigned only to analyzed samples when
using the KBBasecaller. The Score column is blank if:
• Analysis was not performed
• Analysis failed
• ABIBasecaller was used for analysis. The ABIBasecaller does not assign a
score.
5.
Select the checkboxes in the Extract column for the samples to re-extract.
6.
As needed, select different results group and other settings.
7.
Click Extract.
Reextraction creates a new sample file and does not replace the previously saved
sample file. For example, if the first extracted file name is, “sample01.ab1” then
the second first extracted file is assigned “sample01.2.ab1”.
Reextracted sample files are saved in the original folder that data was extracted
to, unless you modify the results group settings.
Chapter 5
Run the instrument
Check the autoextraction status and manually re-extract
5
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
73