9.
To prevent damage to the capillary array, inspect each reservoir assembly and
verify that the following conditions are met.
• Septa fit snugly and flush on the reservoir
• Rubber gasket around the edge of the reservoir cap is seated
1
1
Rubber gasket not seated correctly
• Plate retainer holes and the septa strip are aligned
1
1
Plate retainer holes and septa holes are not aligned
10.
Dry the reservoirs using lint-free wipes.
1
1
Lint-free wipe
11.
Connect the buffer reservoir plate base cable to the heater outlet inside the
instrument.
IMPORTANT!
After installing the buffer reservoir, make sure the cable is out of
the way of the autosampler.
1
2
1
Buffer reservoir
2
Buffer position
12.
Place the water and waste reservoirs in the instrument. Load the three reservoirs
in the following order.
a.
Buffer reservoir
Chapter 9
Maintenance
Perform a short-term shutdown
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
129