Wizards types and functions
To display the Wizard menu at the top of the screen, select GA Instruments
4
ga3730
4
Instrument name.
There are six Wizards available.
Wizard
When to use
Install Array
• To install a capillary array
• To replace an installed capillary array with another capillary array
Change Polymer Type
• To change to a different polymer type
Replenish Polymer
• To replenish the polymer supply
• To replace the polymer in the polymer delivery pump with polymer of the same or
different lot
Bubble Remove
• To remove bubbles in the polymer delivery pump chamber, channels, array
ferrule, and tubing
Water Wash
• To wash the polymer delivery pump chamber, lower polymer block, channels, and
tubing with water when the following conditions are met.
– As part of a monthly maintenance protocol
– To remove any suspected contaminants in the polymer delivery pump
– To remove persistent bubbles (followed by the Bubble Remove wizard, if
needed)
– To replace old polymer in the polymer delivery pump
Instrument Shutdown
• To prepare the instrument for a shutdown >1 week
• To remove the capillary array
Chapter 9
Maintenance
Wizards types and functions
9
116
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide