2.
If there is insufficient space on E: drive, archive the sample files to another
location, then delete the files from E: drive.
Note:
The software requires at least 2 GB of free disk space on E: for proper
operation. Move files from the E:\drive to archival locations regularly.
Note:
Do not specify C: drive locations for data storage. C: drive is used for the
operating system. The software does not perform pre-run disk space checks on C:
drive.
The Cleanup Database function deletes the following content from the database.
Deleted
Not deleted
• Sequence analysis and fragment
analysis plate records with a status of
"Processed"
• All run data in the database that
corresponds to the plate records
•
Spectral calibration plate records
•
Sequence analysis and fragment
analysis plate records with a status of
"Completed"
•
Instrument protocols
•
Result groups
•
Analysis protocols
IMPORTANT!
Do not delete spectral calibration plate records. Doing so permanently
deletes the spectral information from the database and prevents you from re-
extracting data from the plate record.
1.
In the navigation pane of the Data Collection Software, click GA
Instruments
4
Database Manager.
The Database Manager opens.
2.
Click Cleanup Processed Plates.
The Cleanup Database dialog box opens.
Delete data from
the database
Chapter 9
Maintenance
Maintain adequate space for database and sample data storage
9
132
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide