1.
In the navigation pane of the Data Collection Software, select GA
Instruments
4
ga3730
4
Plate Manager.
2.
At the bottom of the screen, click New.
3.
Enter plate record information, then select Spectral Calibration in the
Application dropdown list.
4.
In row A01, enter a sample name, click the next cell, then enter a comment if
needed.
5.
In column 1 of row A01, select the instrument protocol you created for the
spectral calibration.
Create a spectral
calibration plate
Chapter 8
Calibrate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
100
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide