IMPORTANT!
Use only black plate bases with septa-sealed plates. If you are
using MicroAmp
™
Fast 96-Well Reaction Plates (0.1 mL), use only blue plate
bases and matching retainer.
1
2
3
4
1
2
4
5
3
4
1
Plate retainer
2
Heat-sealed sample plate
3
Gray plate base
4
Assembled components
IMPORTANT!
Use only gray plate bases with heat-sealed plates. If you are using
MicroAmp
™
Fast 96-Well Reaction Plates (0.1 mL), use only dark green plate
base and matching retainer.
2.
Verify that the holes of the plate retainer and the septa are aligned.
IMPORTANT!
The plate may damage the array if the retainer and the septum
holes are not aligned.
3.
Make sure when you assemble a plate that the retainer clip is flush with the plate
base. A simple way to ensure that they are flush is to run your finger along the
edge.
Chapter 2
Prepare the instrument
Seal and assemble plates
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
43