3.
Examine each row of the 96 Capillary Position table. Typical values for the Left
spacing and Right spacing columns are:
• 4–8 pixels for a 96-capillary array
• 9–11 pixels for a 48-capillary array
Note:
Higher values are acceptable if a corresponding gap in the capillaries in
the detection cell are visible.
Examine every position for the capillary array.
4.
Click Accept or Reject.
Chapter 8
Calibrate the instrument
Perform spatial calibration
8
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
93