6.
Fill in the remaining rows by manually entering and selecting information, by
using Fill Down commands, or by copy/pasting.
Fill Down
Copy/paste
Select the entire row, select Edit
4
Fill
Down Special, then select a fill down
option.
For 48-capillary instruments, select Fill
down Special (48 Cap).
For 96-capillary instruments:
• 96-well plate: Select Fill Down.
• 384-well plate: Fill down Special
(96 Cap).
• If you copy multiple cells, select the
same number of corresponding
target cells before you paste.
• You can copy/paste within one
protocol.
Note:
Use the Duplicate function in
the Plate Manager to copy
information from an existing
protocol to a new protocol.
7.
Click OK.
1.
In the navigation pane of the Data Collection Software, select GA
Instruments
4
ga3730
4
Instrument name
4
Run Scheduler.
2.
In the Add Plate field, type the Plate ID then press Enter or scan the bar code of
a plate to add it to the input stack.
3.
Place the plate assembly into the instrument. See “Place plate assemblies into the
Run the spectral
calibration plate
Chapter 8
Calibrate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
101