Table 7
Sequencing dye sets and standards
Kit used for sequencing
Dye Set
Calibration Standards
BigDye
™
Terminator v3.1
Cycle Sequencing Kit
Z_BigDyeV3
3730/3730xl DNA Analyzer
Sequencing Standards,
BigDye
™
Terminator v3.1
BigDye
™
Direct Cycle
Sequencing Kit
Z_BigDyeV3
BigDye
™
Terminator v1.1
Cycle Sequencing Kit
E_BigDyeV1
3730/3730xl DNA Analyzer
Sequencing Standards,
BigDye
™
Terminator v1.1
Table 8
Fragment analysis dye sets and matrix standards
Dye Set
Calibration Standards
G5
DS-33
G5-RCT (reduced cross-talk)
[1]
J6
DS-36
Any4Dye, Any5Dye
Custom
Any4Dye-HDR (high dynamic range)
[2,
3]
Custom
Any5Dye
DS-02 for SNaPshot
™
applications
Custom
[1]
Use for fragment analysis applications on 96-capillary arrays.
[2]
Signal intensity is reduced by approximately half relative to the standard dye sets, along with a minimal
reduction in the noise, resulting in a very slight decrease in the signal/noise ratio when compared to data
generated using the standard dye sets
[3]
When using this dye set, perform spectral calibrations each time the capillary array is replaced or moved
within the detection cell
• 384- or 96-Well Reaction Plate w/Barcode
• Multichannel pipettor
• Plate retainer
– Plate septum with black plate base
or
– Heat-seal with gray plate base
• Hi-Di
™
Formamide
• Heated block or thermal cycler
• Container with ice
• Centrifuge with microplate adapter
• Microcentrifuge
• Vortex
• Gloves
Supported dye
sets and
calibration
standards
Required
materials
Chapter 8
Calibrate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
98
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide