Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 4
Calibration verification test
Evaluating the
results
When the GeneAmp
®
9700 completes the calibration verification test, one of two
screens opens. See the table below to evaluate the results.
Ending the test
1. Press
F5
(Exit).
2. Remove the 0.2-mL probe assembly from the sample block.
3. Power off the digital thermometer, then clean off the oil.
4. Wait for the sample block to reach room temperature (~25 °C), then remove the
dual 96-well temperature verification plates from the sample block.
PHYSICAL INJURY HAZARD.
Hot Surface. Use care
when working around the heated cover and sample block to avoid being burned
by hot components.
5. Clean the oil off the sample block.
If the sample block
module...
Then the...
Is properly calibrated
Calibration Verification screen displays the following message:
Does not pass the
calibration verification
test
Calibration Verification screen displays the following message:
• If the test fails, repeat the procedure to make sure that the
thermometer was not misread or that data-entry errors
were not made.
• If the test fails again, contact Applied Biosystems Technical
Support. See
“How to obtain support” on page viii
.
Calibration Verification
Calibration is good
F1
F3
F4
F5
F2
Exit
Calibration Verification
Instrument may require service.
Contact Applied Biosystems
F1
F3
F4
F5
F2
Exit
Technical Support.