120
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 10
Calibration verification test
5. Place the 0.5-mL Thermal Isolation Frame onto the sample block.
6. Using a cotton swab, coat the following wells with mineral oil:
B3, B8, F1, F10
7. Place the 0.5-mL probe assembly into well B3. Thread the probe assembly wire
through the channel in the temperature verification frame to prevent damage to
the probe assembly and lead wires (see
).
IMPORTANT!
Seat the probe properly and the close the heated cover carefully.
If the probe wire is crushed when you close the heated cover, the probe can be
damaged.
8. Power on the digital thermometer.
Refer to the instructions in your Temperature Verification Kit for details on
operating the digital thermometer, Model 4500.
9. Make sure the probe assembly is connected to the digital thermometer and that
the thermometer is powered on.
10. Slide the heated cover forward and pull the lever down.