Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 3
Overview
Overview
Materials
When you perform a temperature verification test on a GeneAmp
®
9700 with a 96-
Well Sample Block Module, use the materials in the kit (PN 4317939).
Assembly
shows a properly assembled system. Note that the frame is positioned
around the wells with the channel facing you. The location tab on the frame should
align with the locator tab on the thermal cycler. The probe wire is threaded through
the channel to prevent damaging the wire when you close the thermal cycler heated
cover.
Figure 5
Proper assembly of the Temperature Verification System for
GeneAmp
®
9700 thermal cycler
Figure 6
Temperature verification frame