Digestion Times
When you press the
Digestion Times
button the following screen will be displayed.
This diagnostic allows you to modify the times used by the computer for the digestion processes. This can be helpful when
troubleshooting or demonstrating the instrument. If you press the
Save to Permanent Memory
button, the values you entered
will be saved until you change them again by using this diagnostic. If you do not press the
Save to Permanent Memory
button,
the values you entered will be saved until power is turned off, or until you change them again by using this diagnostic.
If you start a new procedure with Digestion Times other than the official method
times, a message will be displayed on the Touch Screen Display that allows you to
reset the values. A Reset from this Warning Screen is only temporary. Use the
Digestion Times screen in Diagnostics to make the reset permanent.
Pump Tube Test
When you press the
Pump Tube Test
button on the Diagnostics screen, the following screen will be displayed.
This diagnostic tests the integrity of the tubing from the pump to the inlet valves (IDF, SDF, and Waste). When you press the
START
button, the following screen will be displayed.
All lines should be charged before test is run.
Verify that all supply containers are connected and contain
the desired fluid.
Cancel
Continue
(no line charge)
Continue
(line charge)
Pump Tube Test
START
Status
Pressure
difference
(psi)
1
2
3
4
5
6
Vacuum xx.x
cur: - x.x
max: - x.x
1.2 ml delivery
←
AOAC 991.43/985.29/2001.03
Digestion Times
(Press button with number to change)
Amylase
Digestion Time
991.43/985.29 SDF
Precipitation Time
Protease
Digestion Time
HCl/AMG
Digestion Time
Save to
Permanent
Memory
←
(minutes)
20
30
(minutes)
30
(minutes)
(minutes)
60
Reset to official
method times
(hours)
12
2001.03 SDF
Precipitation
Содержание HS
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 44: ...Operator s Manual pg 44 Rev 9 20 18 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev 9 20 18 This page intentionally left blank ...
Страница 80: ...Operator s Manual pg 80 Rev 9 20 18 This page intentionally left blank ...