Operator’s Manual
Rev 9/20/18
pg. 53
15.
Verify that the fluid containers have the correct fluids for this procedure.
Verify that the fluid supply containers are configured according to the screen above and press the check mark (
) button.
One of the following screens will be displayed.
16.
Fill fluid containers.
To ensure that you have enough fluids to run a complete TDF procedure, you must begin with fluid levels above the
Minimum Level lines on the chemical containers and at least 15 ml of each enzyme. Add fluids and enzymes as
necessary.
Do NOT leave the enzyme ports on the instrument open to the air or the enzymes
in the valves may dry up and plug the ports.
With all fluid containers filled to the proper levels, press the check mark (
) button on the Touch Screen Display. One of
the following screens will be displayed.
AOAC
2001.03
TDF
←
Nitrogen supply ON?
AOAC
985.29
TDF
←
Nitrogen supply ON?
←
Fluid levels correct?
AOAC
2001.03
TDF
←
Fluid levels correct?
AOAC
985.29
TDF
AOAC
2001.03
TDF
←
Check supply containers
Amylase
AMG
HCL
Buffer
-> Amylase
-> AMG
-> HCL
-> Phosphate Buffer
AOAC
985.29
TDF
←
Check supply containers
Amylase
AMG
HCL
Buffer
-> Amylase
-> AMG
-> HCL
-> Phosphate Buffer
Содержание HS
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 44: ...Operator s Manual pg 44 Rev 9 20 18 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev 9 20 18 This page intentionally left blank ...
Страница 80: ...Operator s Manual pg 80 Rev 9 20 18 This page intentionally left blank ...