Operator’s Manual
Rev 9/20/18
pg. 45
TDF Analysis (AOAC 985.29 / 2001.03)
A TDF analysis directly measures the amount of TDF within a given sample without separately measuring the IDF and SDF
fractions. TDF can be determined using multiple methods. This section describes the procedure for using the ANKOM
TDF
Dietary Fiber Analyzer to determine TDF based on the AOAC 985.29 and 2001.03 methods.
This analysis requires an SDF filter bag (ANKOM DF-S) for the precipitation process and a non-filter IDF
Flow-Thru
bag
(ANKOM DF-FT) for the digestion process. When starting a new run, the ANKOM
TDF
Dietary Fiber Analyzer must have bags
installed at all stations (a total of twelve bags).
To enhance the productivity of your instrument, you can begin the IDF process of a
new run while the SDF process of a previous run is finishing. See the “Productivity
Enhancement” section of this manual for more details.
To perform a TDF analysis, follow the steps below.
1.
Prepare chemicals and enzymes.
When using the ANKOM
TDF
Dietary Fiber Analyzer for the AOAC 985.29 or 2001.03 methods use the chemicals and
enzymes referred to therein. See Appendix A of this manual for the list of chemicals and enzymes and the instructions for
how to prepare them for use in this instrument.
2.
De-fat your samples as needed according to the official methods. See "Appendix C
–
De-fatting
Procedure" in this manual for an ANKOM procedure.
3.
Label the bags using a Solvent Resistant Marker.
4.
Prepare for data collection.
You will need a place to store the data collected during this analysis. For your convenience a CD titled “Calculation
Templates” is included with the instrument. This CD has TDF Data Spreadsheets that can be used for this analysis. Please
read the “Instructions” tab in the MS-Excel file for information about the spreadsheets.
5.
Weigh filter bags.
Using a Bag Weigh Holder is critical to eliminate the effects of static electricity
during the weighing process.
Roll or fold each bag and place it in a tared Bag Weigh Holder. Place the Bag Weigh Holder in the center of a balance
and record the weight.
Because different balances have different sensitivities, the Bag Weigh Holder
should be placed in the center of the balance for best results.
SDF Bag (longer bag with filter)
IDF
Flow-thru
Bag (shorter bag / no filter)
Содержание HS
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 44: ...Operator s Manual pg 44 Rev 9 20 18 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev 9 20 18 This page intentionally left blank ...
Страница 80: ...Operator s Manual pg 80 Rev 9 20 18 This page intentionally left blank ...