Operator’s Manual
Rev 9/20/18
pg. 21
10.4
With the bag being held by the Delivery Nozzle, center it horizontally between the lines on the back part of
Clamp Bar C.
10.5
Repeat steps 10.2 – 10.4 for all six stations.
10.6
Re-install Clamp Bar D by setting it on the locator rods. Make sure the letter is on the top of the bar and the
rubber material is facing in toward the instrument.
10.7
Flatten the bags to remove any wrinkles.
10.8
With fingers away from the clamp bars, press the check mark (
) button on the Touch Screen Display to pinch
the bags just above the filter. The following screen will be displayed.
11.
Add DE to the SDF filter bags on the ANKOM
TDF
Dietary Fiber Analyzer.
11.1
Open the top of the SDF bag and add a weigh tin of DE to the bag by folding the tin and then dipping it down into
the bag below the tip of the Delivery Nozzle.
When adding DE to the filter bags it is very important to keep it below the tip of
the Delivery Nozzle so that the DE material can be properly rinsed.
11.2
Rinse the tin with no more than 3 ml of 78% EtOH to ensure complete transfer.
11.3
Repeat steps 11.1 – 11.2 for all six stations.
Diatomaceous Earth
added in SDF Bags?
AOAC
991.43
IDF/SDF
←
Centering Lines