Operator’s Manual
pg. 16
Rev 9/20/18
Dietary Fiber Analysis Support Items
The following support items needed to perform the fiber analysis may be purchased separately:
Item
ANKOM Part #
Electronic Balance with four-place readout
TB
Electronic Balance Software
TBS
Heat Sealer for sealing the filter bags
HS (120V), HSi (220V)
IDF Filter Bags
DF-I
SDF Filter Bags (used for SDF and TDF procedures)
DF-S
IDF
Flow-Thru
Bags (TDF procedure only)
DF-FT
Solvent Resistant Marker
F06
Rinse Stand
TDF51
Diatomaceous Earth (DE)
DE1, DE2
Drying Oven
Ashing Oven
Protein Determination Equipment (Kjeldahl is recommended)
Analysis Options using the ANKOM
TDF
Dietary Fiber Analyzer
The ANKOM
TDF
Dietary Fiber Analyzer can be configured to recover IDF and SDF residue, or to recover TDF residue. The
following describes these analyses along with their specific filter bag configurations.
Analysis Name
Analysis Description
Upper Bag
Lower Bag
IDF/SDF
(991.43)
This analysis is used to determine IDF and SDF residue separately
using two different filter bags. TDF is calculated by adding the IDF
and SDF residue weights corrected for protein and ash content.
IDF
SDF
TDF
(991.43, 985.29 &
2001.03)
This analysis is used to determine TDF residue in one filter. In this
analysis, the TDF value consists of the weighed residue corrected
for protein and ash.
IDF
Flow-Thru
(no filter)
SDF
A computer controlled Diagnostics mode is provided through the Control Panel for maintenance and troubleshooting purposes.
To maximize productivity, this instrument also allows a user to start new procedures (IDF/SDF and TDF) before the previous
ones have completed. See the “Productivity Enhancement” section of this manual for details.
Contact ANKOM about an upgrade to Total Integrated Dietary Fiber (AOAC
2009.01 and 2011.25 methods). See the Diagnostics section of this document for
instructions on how to enable these methods.
Blanks
BLANK values are used in the IDF, SDF, and TDF calculations. The following BLANK values (based on extensive research
done in the ANKOM lab) are the default values loaded in the IDF_SDF and TDF spreadsheets in the MS-Excel workbook
provided with the instrument.
IDF
(AOAC 991.43 method):
-0.0072
TDF
(AOAC 991.43 method):
-0.0047
SDF
(AOAC 991.43 method):
-0.0030
TDF
(AOAC 985.29 method):
-0.0028
You do not need to run BLANKS with every set of samples. However, if you want to determine your own BLANK values,
especially when you change chemical lots,
ANKOM recommends that the BLANK value in the IDF, SDF, and TDF
calculations be the average of at least 18 BLANK values.
If you determine your own BLANK values, you can overwrite the values currently in the MS-Excel spreadsheets.
Содержание HS
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 44: ...Operator s Manual pg 44 Rev 9 20 18 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev 9 20 18 This page intentionally left blank ...
Страница 80: ...Operator s Manual pg 80 Rev 9 20 18 This page intentionally left blank ...