Operator’s Manual
Rev 9/20/18
pg. 75
Motor Test
When you press the
Motor Test
button on the Diagnostics screen, the following screen will be displayed.
This diagnostic allows you to test the motor at different speeds (default is 240 RPM) and directions using different amounts of
fluid. Default valve settings are Water Supply open and Waste Output open. If you want to change valve settings, before
pressing the
GO
button to start the Motor Test, press the
Set Valves
button. The following screen will be displayed.
This diagnostic also displays the maximum pressure read from the vacuum sensor during the test. The value starts at zero and
will display a value while the motor is running and when it stops. This can be used for troubleshooting.
Running the Motor Test could leave fluid in some of the lines. After running a
Motor Test, always run the Clear IDF/SDF Lines diagnostic.
Set Valves
Amylase Supply
(closed)
HCl Supply
(closed)
AMG Supply
(closed)
Protease Supply
(closed)
Water Supply
(closed)
EtOH95 Supply
(closed)
EtOH78 Supply
(closed)
Buffer Supply
(closed)
IDF Inlet
(open)
Waste
(open)
SDF Inlet
(open)
Output
Supply
Motor Test
Set Valves
GO
Motor
Direction
(FWD)
120 RPM
240 RPM
Motor Speed
10.0
Amount (ml)
Vacuum xx.x
cur: xx.x
max: xx.x
←
←
Содержание HS
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 44: ...Operator s Manual pg 44 Rev 9 20 18 This page intentionally left blank ...
Страница 70: ...Operator s Manual pg 70 Rev 9 20 18 This page intentionally left blank ...
Страница 80: ...Operator s Manual pg 80 Rev 9 20 18 This page intentionally left blank ...