Transporting the contrAA 600
74
Edition 01/2014
contrAA 600
8
Transporting the contrAA 600
CAUTION! RISK OF INJURY IF THE DEVICE FALLS!
Handles which are screwed in too loosely may give rise to damage during transport. Screw
in the handles up to the end stop.
WARNING!
The device must be transported by at least 4 persons using the fixed screw-in carrying
handles (included in scope of supply).
The contrAA 600 is too heavy for two persons (
→
Section "Data on contrAA 600" p. 13). In
addition, it is not possible to get an adequate grip for transport without the screw-in handles.
There is a risk of injury if transport is attempted without using the handles or by too few
people.
Uninstall all components, see Chapter "Installation and start-up" p. 36. Ensure that the outlet
tube has been removed from the sample chamber.
1.
Close the gas supply upstream of the device connections.
2.
Remove the argon tube from the coupling.
3.
Undo the electrical connections.
4.
Remove the four stoppers from the holes for the handles on both sides of the device
and keep in a safe place.
5.
Screw the four handles securely into the holes as far as the end stops.