Installation conditions
contrAA 600
Edition 01/2014
19
A separate room is recommended for preparing samples and storing wet-chemical
materials.
There is a no smoking policy in the room where the contrAA 600 is operated.
Temperature range during ope10 °C to 40 °C
Temperature range during storage -40 °C to 70 °C, use drying agent
Humidity during operation
max. 93 % at 40 °C
Humidity during storage
10% to 30%, use drying agent
4.2
Space requirement and weight
Minimum size of work table:
1880 mm
×
700 mm, select the height according to ergonomic requirements
Carrying capacity of work table: at least 200 kg
Table surfaces: Wipe, scrape and corrosion resistant, not allowed to absorb moisture
Set up the work table in such a way to allow easy access from all sides.
Components
Width [mm]
Height [mm]
Depth [mm]
Weight [kg]
On the work table
contrAA 600
785
653
765(974)
160
AS-GF
250
550
380
7,2
AS-F
340
350
460
6,5
AS-FD
Autosampler
Fluidik module
340
360
350
310
460
165
6,5
3,5
HS 60 modular
360
370
240
14
HS 55 modular
360
370
240
14
Table 1
Measurements and weights of the components of the contrAA 600
4.3
Energy supply
OBSERVE THE MAINS CONNECTION!
During electrical installation observe the VDE (German Association for Electrical Engineers)
guidelines and local regulation requirements!
The mains supply must be correctly earthed.
Do not use an adapter in the mains cabling.
Do not use an automatic fuse device.
The contrAA 600 is operated on single-phase alternating current. The current load can reach
85 A for a short period (1 s) during maximum heating. The mains voltage at the contrAA 600
should not decrease by more than 6 % during this period. For any deviation from these
values, please contact Analytik Jena AG. Appropriate accessories can be supplied.