Specification
contrAA 600
Edition 01/2014
17
Autosampler AS-FD
Autosampler with dilution function, completely PC-controlled
Sample tray 139/ 15
Sample cups
Special cups
129 pieces, 15 mL
10 pieces, 50 mL
Sample tray 54/ 50
Sample cups
54 pieces, 50 mL
Dosing unit in the Fluidik module
5 mL
Power supply
Via AAS basic instrument
Wash bottle
2 L
Bottle for diluent
2 L
Mass (total)
Autosampler
Fluidik module
10.0 kg
6.5 kg
3.5 kg
3.2
Guidelines and standards
Safety class and safety type
The contrAA 600 belongs to safety class I.
The casing is safety type IP 20.
Device safety
The contrAA 600 conforms to the safety standards
DIN EN 61010-1 (VDE 0411T.1; IEC 61010-1)
DIN EN 61010-2-061 (IEC 61010-2-061)
EMC compatibility
The contrAA 600 has been tested for radio interference elimination and interference immuni-
ty and fulfills the requirements stipulated by
DIN EN 61326
Environmental compatibility
The contrAA 600 has been tested for environmental compatibility and fulfills the require-
ments stipulated by
DIN ISO 9022-3:2000
DIN ISO 9022-32-03-0
DIN ISO 9022-2:2003/01
EC directives
The contrAA 600 is built and tested according to standards that fulfill the requirements
stipulated by the EC directives 2006/95/EC and 2004/108/EC. Each device leaves the manu-
facturer in a pristine and technically safe state. To maintain this condition and to ensure safe
operation, the operator must strictly observe the safety and operating instructions contained
in this manual. For accessories which have also been supplied, and system components