Installation and start-up
52
Edition 01/2014
contrAA 600
Preparing the Fluidik module (for AS-FD)
1
Storage bottle for washing liquid
2
Diluent connection
3
Dosing tube connection (to AS-FD)
4
Dosing syringe, consisting of piston
and glass cylinder
5
Dosing syringe with attachment screw
6
Storage bottle for diluent
Fig. 33
Dosing unit at the Fluidik module of the AS-FD
1.
If necessary, fit the dosing syringe to the dosing unit (
→
Section "Replacing the meter-
2.
Place the storage bottles for the wash liquid (left) and diluent (right) into the bottle
holders of the Fluidik module.
3.
Immerse the short tube (marking at the tube "3") into the storage bottle for the diluent.
Screw the second tube end to the valve (2 in Fig. 33).
4.
Screw the dosing tube for the diluent (encased, marking "1") to the second connection
of the valve (3 in Fig. 33).
5.
Immerse the hose for the wash liquid (marking "2") into the storage bottle.
Work steps: Uninstall
1.
Switch off the contrAA 600.
2. Pull off the sample tube from the aspirating canula on the autosampler arm.
3. Pull off the outlet tube from the connector of the autosampler (rearside).
4. Loosen the control cable on the rear of the autosampler.
6.8
Starting up the contrAA 600 with accessories
Switch-on sequence, daily work commencement
1.
Switch on the PC and wait for the computer program to initialize: The application icons
appear on the screen, including the ASpect CS program icon.
2.
Switch on the contrAA 600: Press the green ON/OFF switch on the right side wall.
3.
Start the ASpect CS program: Double-click with the mouse cursor on the ASpect CS
icon.