6 — Troubleshooting
34
11830378_C
AMSCO
®
Rack Return
Operator Manual
General Troubleshooting
Additional Information
• If status indicator lighting is present, lights are red
on both sides.
indicator lighting is present: opened door light
turns white or green.
3. Press Command button on opposite side to close
door and return to normal operation.
Hands-free sensor is malfunctioning.
• If status indicator lighting is present, light is green or
white.
1. Press Command button to open door.
2. If door opens, try cleaning hands-free sensor. If
hands-free sensor is still not working, call STERIS
(service charges may be incurred).
3. If door does not open, restart rack return (refer to
section relating to
System Software is Malfunctioning
in
Operator Manual
).
Command button is malfunctioning.
• If status indicator lighting is present, light is green or
white.
1. Obstruct hands-free sensor to open door.
2. If door opens, Command button is malfunctioning.
Call STERIS (service charges may be incurred).
System software is malfunctioning.
• If status indicator lighting is present, light is green or
white.
1. Turn rack return power OFF.
2. Turn rack return power ON.
3. If problem reoccurs, call STERIS (service charges
may be incurred).
Door stopped moving before being completely open or
closed.
• If status indicator lighting is present, light is red on
one side.
1. Verify if something is obstructing door.
2. If so, remove obstruction.
3. Press Command button.
Pressure is too low.
1. Verify if there is air pressure leakage.
2. If so, call STERIS (service charges may be incurred).
One of the door cylinders is defective.
1. Call STERIS (service charges may be incurred).
Emergency stop pushbutton has been pressed.
1. Resolve situation.
2. Pull Emergency stop pushbutton.
3. Press Command button to acknowledge alarm and
resume normal operation. NOTE: If SCS Conveyor
System or ATLAS WAV System is present and doors
are opened, press Command button to close all
doors.
Emergency stop pushbutton is malfunctioning.
1. Pull Emergency stop pushbutton.
2. Press Command button to acknowledge alarm.
3. If door does not open, Emergency stop pushbutton
is malfunctioning. Call STERIS (Service charges may
be incurred).
Содержание STERIS
Страница 4: ...iv 11830378_C AMSCO Rack Return Operator Manual ...
Страница 6: ...vi 11830378_C AMSCO Rack Return Operator Manual ...
Страница 8: ...viii 11830378_C AMSCO Rack Return Operator Manual ...
Страница 10: ...x 11830378_C AMSCO Rack Return Operator Manual ...
Страница 14: ...4 11830378_C AMSCO Rack Return Operator Manual ...
Страница 18: ...8 11830378_C AMSCO Rack Return Operator Manual ...
Страница 28: ...18 11830378_C AMSCO Rack Return Operator Manual ...
Страница 34: ...24 11830378_C AMSCO Rack Return Operator Manual ...
Страница 40: ...30 11830378_C AMSCO Rack Return Operator Manual ...
Страница 47: ......
Страница 48: ......