6 — Troubleshooting
33
Operator Manual
AMSCO
®
Rack Return
11830378_C
General Troubleshooting
Additional Information
Command button is malfunctioning.
• If status indicator lighting is present, light is green or
white.
1. Call STERIS (service charges may be incurred).
Door took too long to close, sensor was not activated
within required time.
• If status indicator lighting is present, light is red on
one side.
1. Verify if something is obstructing door. If so,
remove obstruction. Press Command button.
2. Pressure is too low. Verify if there is air pressure
leakage. Call STERIS (service charges may be
incurred).
3. Bottom sensor is malfunctioning. Call STERIS
(service charges may be incurred).
Door is obstructed.
• If status indicator lighting is present, light is red on
one side.
1. Remove obstruction and press Command button.
2. Obstruction sensor is malfunctioning. Call STERIS
(service charges may be incurred).
Command button was pressed too quickly after the door
was open.
• If status indicator lighting is present, light is red on
one side.
1. Press Command button to acknowledge alarm. If
optional status indicator lighting is present: light
turns green or white. Wait a few seconds before
pressing Command button once again.
2. Press Command button to close door.
Rack sensor is malfunctioning.
• If status indicator lighting is present, light is green or
white.
1. Press Command button.
2. Push rack until it rests against the bumpers.
3. Press Command button to close door and return to
automatic mode.
4. If door reopens, rack sensor is malfunctioning. Call
STERIS (Service charges may be incurred).
Wrong hand-free sensor sensibility settings.
1. Call STERIS (Service charges may be incurred).
6.5 Door Does Not Open
Condition
Door does not open when Command button is pressed or hands-free sensor is obstructed.
General Troubleshooting
Additional Information
Door interlock is activated.
1. Door does not open until door on opposite side is
closed.
Both doors are open, preventing rack return to function
normally. Verify if something is preventing door from
moving.
1. Press Command button to acknowledge alarm and
open door.
2. Press Command button to close door. Opposite
door automatically opens. If optional status
Содержание STERIS
Страница 4: ...iv 11830378_C AMSCO Rack Return Operator Manual ...
Страница 6: ...vi 11830378_C AMSCO Rack Return Operator Manual ...
Страница 8: ...viii 11830378_C AMSCO Rack Return Operator Manual ...
Страница 10: ...x 11830378_C AMSCO Rack Return Operator Manual ...
Страница 14: ...4 11830378_C AMSCO Rack Return Operator Manual ...
Страница 18: ...8 11830378_C AMSCO Rack Return Operator Manual ...
Страница 28: ...18 11830378_C AMSCO Rack Return Operator Manual ...
Страница 34: ...24 11830378_C AMSCO Rack Return Operator Manual ...
Страница 40: ...30 11830378_C AMSCO Rack Return Operator Manual ...
Страница 47: ......
Страница 48: ......