4 — Operating Instructions
20
11830378_C
AMSCO
®
Rack Return
Operator Manual
WARNING
SLIPPING HAZARD
To prevent slips, keep floors dry. Promptly clean up any spills or drippage. For spills or
drippage of detergents or other chemicals, follow safety precautions and handling
procedures set forth on detergent or chemical label and/or Safety Data Sheet (SDS).
4.3 Before Operating Equipment
WARNING
PERSONAL INJURY
Risk of pinch point between door and upper panel. Do not push on top portion of doors;
do not push on door when door is rising; do not push on door when something is
jammed.
Risk of pinch point between door and threshold when door closes. Keep fingers away
from threshold.
1. Verify POWER (ON/OFF) toggle switches are positioned to ON.
4.4 Manually Changing Indicator Lighting Status
NOTE:
This procedure only applies with the status indicator lighting accessory installed.
The rack return may be equipped with status indicator lighting that informs the operator about the rack return status.
In some situations, the indicator lighting status may not represent the actual rack return status. To manually change
the lighting status, ensure the opposite door is closed and press and hold the
Command
button for three seconds.
NOTE:
This function does not work when the rack return is in alarm state.
The status indicator lighting indicates the following status:
• Green lights: Chamber is available for use.
• White light: Chamber is ready to be unloaded.
• Red light: An alarm was triggered.
• Light off: Chamber is not available for use.
4.5 Door Operation
The rack return is equipped with two sliding doors that can be opened using the
Command
button or with the hands-
free sensor. Doors are closed using the
Command
button. In some situations, doors may be opened unintentionally. In
that case, press the
Command
button to close door.
If rack return is operated automatically with ATLAS WAV System or SCS Conveyor System
and door is opened
manually using the
Command
button, rack return will stop working automatically. To resume automated operations,
press
Command
button to close door.
4.6 Loading Rack Return Manually
1. Ensure that the status indicator lighting (accessory) is green on rack return.
Содержание STERIS
Страница 4: ...iv 11830378_C AMSCO Rack Return Operator Manual ...
Страница 6: ...vi 11830378_C AMSCO Rack Return Operator Manual ...
Страница 8: ...viii 11830378_C AMSCO Rack Return Operator Manual ...
Страница 10: ...x 11830378_C AMSCO Rack Return Operator Manual ...
Страница 14: ...4 11830378_C AMSCO Rack Return Operator Manual ...
Страница 18: ...8 11830378_C AMSCO Rack Return Operator Manual ...
Страница 28: ...18 11830378_C AMSCO Rack Return Operator Manual ...
Страница 34: ...24 11830378_C AMSCO Rack Return Operator Manual ...
Страница 40: ...30 11830378_C AMSCO Rack Return Operator Manual ...
Страница 47: ......
Страница 48: ......