6 — Troubleshooting
32
11830378_C
AMSCO
®
Rack Return
Operator Manual
WARNING
SLIPPING HAZARD
To prevent slips, keep floors dry. Promptly clean up any spills or drippage of detergents
or other chemicals, follow safety precautions and handling procedures set forth on
detergent or chemical label and/or Safety Data Sheet.
6.2 Abnormal Noise Coming From Rack Return
Condition
Abnormal noise is heard during rack return normal operation.
General Troubleshooting
Additional Information
A hose is disconnected.
1. Call STERIS (service charges may be incurred).
An object might be stuck in chamber.
1. Verify chamber. If so, remove object.
6.3 Command Button Does Not Work
Condition
Command button does not work.
General Troubleshooting
Additional Information
Opposite door is not closed.
1. Verify if opposite door is closed.
Equipment POWER (ON/OFF) Toggle Switch is OFF.
1. Verify both POWER (ON/OFF) switches are
positioned to ON.
There is no power to equipment.
1. Plug equipment to a power source.
Command button is damaged.
1. Call STERIS (services charges may be incurred).
6.4 Door Does Not Close
Condition
Door does not close when Command button is pressed.
Содержание STERIS
Страница 4: ...iv 11830378_C AMSCO Rack Return Operator Manual ...
Страница 6: ...vi 11830378_C AMSCO Rack Return Operator Manual ...
Страница 8: ...viii 11830378_C AMSCO Rack Return Operator Manual ...
Страница 10: ...x 11830378_C AMSCO Rack Return Operator Manual ...
Страница 14: ...4 11830378_C AMSCO Rack Return Operator Manual ...
Страница 18: ...8 11830378_C AMSCO Rack Return Operator Manual ...
Страница 28: ...18 11830378_C AMSCO Rack Return Operator Manual ...
Страница 34: ...24 11830378_C AMSCO Rack Return Operator Manual ...
Страница 40: ...30 11830378_C AMSCO Rack Return Operator Manual ...
Страница 47: ......
Страница 48: ......