5-2
129359-452, Volume 1
Service and Maintenance Procedures
Operator Troubleshooting
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
9. Jacket pressure does not
Steam supply hand valve not open — open valve.
rise to required level.
Possible flood condition/steam supply solenoid valve closed.*
Steam supply less than 50 psig (3.52 kg/sq.cm) dynamic — check
supply pressure and increase if necessary.
10. Cycle does not advance from
Door gasket worn or defective — replace.
prevacuum phase.
Chamber drain line or strainer plugged — clean line or strainer.
Water supply hand valve not open — open valve.
Water supply less than 30 psig (2.11 kg/sq.cm) dynamic — check
supply pressure and increase if necessary.
Steam supply insufficient — see 9 above; “Jacket pressure does
not rise . . .”
HI-LO valve not set correctly — check that it is set on HI.