3-6
129359-452, Volume 1
Service and Maintenance Procedures
Service Adjustments and Special Features
c. Increase or decrease the value (of blinking digit) by pressing the VALUE
direction pad (up or down).
d. Repeat steps b. and c. until correct time and date is set.
e. Press SAVE VALUES pad to record the new time and date and to advance
control to ready mode. Printout of cycle values and new time and date will
also be furnished.
3.7
Set Automatic Utilities Control Times (if
feature activated)
NOTE: When Automatic Utilities Control feature is set to OFF (service adjust-
ment), no printouts or displays will indicate that the feature exists.
Sterilizer can be set to automatically control utilities shutdown and restart
times. The control will automatically shut off utilities when the SHUTDOWN set
time is reached and turn them back on when the RESTART set time is reached.
If a cycle is in progress when the SHUTDOWN time is reached, the utilities
will not be shut off until the cycle is complete.
The following conditions apply when the Automatic Utilities Control feature is
in use:
• If the SHUTDOWN and RESTART times are identical or if SHUTDOWN time
is set to zero, then no shutdown will occur.
• If the RESTART time is set to zero, then no restart will occur until a CYCLE
touch pad (numbered
1, 2, 3
and
4
) or RESET touch pad is pressed.
• RESTART can be activated manually by pressing a CYCLE touch pad
(numbered
1, 2, 3
and
4
) or pressing RESET touch pad when in the
SHUTDOWN phase.
• If a power failure occurs during SHUTDOWN, the control will return to
SHUTDOWN phase upon return of power.
When the control is in the SHUTDOWN phase the display will show:
UTILITIES SHUTDOWN
RESTART AT (SET TIME)
1. Open the printer door and position the power switch to ON.
• Display panel lights up.
• The printer records the time and date that the power is turned ON.
NOTE: When power to prevacuum sterilizers is switched on, display will show
the following messages:
•
FIRST MESSAGE
: PERFORM LEAK TEST? Press cycle selector touch
pad number 2 for NO.
•
SECOND MESSAGE
: PERFORM DART TEST? (BOWIE-DICK) Press
cycle selector touch pad number two for NO.
2. Close the printer door and raise the sliding door over the touch pad panel
to access the VALUES touch pads.
NOTE: If access code feature is enabled, bypass code by pressing CHANGE
VALUES touch pad when NO is blinking.
Figure 3-3. Turn Power On
Содержание EAGLE SERIES 3011