4-2
129359-452, Volume 1
Service and Maintenance Procedures
Routine Maintenance
4.1
Daily (See Figure 4-1)
1. Clean chamber as follows:
a. Wash inside of chamber and loading equipment with a mild detergent
solution such as STERISs’ Liqui-Jet* or Sonic* detergents.
b. Rinse with tap water; dry with a lint-free cloth.
c. Remove chamber drain strainer. Clean out lint and sediment; reverse
flush under running water.
d. Place strainer back in chamber drain.
NOTE: Flush drain whenever line becomes clogged or spillage occurs during
a liquids processing cycle (see Paragraph 4.2).
NOTE: To keep your sterilizer and loading equipment looking like new, STERIS
recommends an occasional application of Pry Cream* to the non-painted
surfaces.
4.2
Weekly (See Figure 4-1)
1. Flush chamber drain as follows:
a. Remove chamber drain strainer.
b. Rinse drain with hot solution of trisodium phosphate (two tablespoons to
one quart of water).
or
Use a solution of 1/2 cup of STERIS Sonic Detergent* and one quart of hot
water if trisodium phosphate is objectionable.
c. Wait five minutes.
d. Flush drain with one quart of hot water.
e. Place strainer back in chamber drain.
2. Check control and status signals as follows:
a. Empty chamber and set cycle values for a one minute gravity cycle.
b. Start cycle and observe gauges, display, and printed tape for proper
functioning. (Observe non-operating end control panel also on double
door models.)
3. Flush steam generator (if supplied with sterilizer) as follows:
a. Position POWER switch to OFF.
b. Open lower access door.
c. Wait until generator pressure gauge indicates zero, then open generator
drain valve by turning valve handle counterclockwise 90 degrees.
d. Position POWER switch to ON and flush generator for five minutes.
e. Close drain valve. The generator will refill with clean water.
f. Position POWER switch to OFF.
g. Close lower access door.
*Available from your local STERIS repre-
sentative.
Содержание EAGLE SERIES 3011
Страница 3: ...Intentionally Blank Page ...