3-1
Service Adjustments and Special Features
Service and Maintenance Procedures
129359-452, Volume 1
SERVICE ADJUSTMENTS AND
SPECIAL FEATURES
3
BURN AND SHOCK HAZ-
ARD: Repairs and adjust-
ments should be attempted
only by authorized persons
fully acquainted with this
equipment. Use of inexperi-
enced, unqualified persons
to work on the equipment or
the installation of unautho-
rized parts could cause per-
sonal injury or result in
costly damage.
Factory settings of certain sterilizer control functions are service adjustable to
best adapt sterilizer to your operation. See Tables 3-1 and 3-2 for a listing of
these features.
NOTE: These adjustments must be made by a qualified service technician.
Procedures for making these adjustments are included in the MAINTENANCE
MANUAL (see Section 7 for ordering information).
TABLE 3-1. SERVICE ADJUSTMENTS
(Dip Switch)
Function
Factory Setting
Optional
Temperature Display
and Printout Units*
Degrees F
Degrees C
Pressure Display
and Printout Units
PSIG
Bar
Time Display and
Printout Units
Std. AM/PM
Military
Automatic Duplicate
Printout
No
Yes
Automatic Utilities
Control**
No
Yes
Intermittent Buzzer
for Cycle Complete
Yes
No
Access Code
For Cycle Values
No
Yes
Point of Use
Laboratory or Hospital
Hospital
Lab***
Undertemperature
Restart
Resume
Recovery
Timer
Timer***
* For both Fahrenheit and Centigrade temperatures (in hospital or laboratory mode),
temperature is set and displayed to the nearest degree; however, temperature
control and printout is to the nearest 1/10 degree.
**Requires optional Utilities Shutdown Valve Kit.
***These adjustments are only for use in a laboratory. See VOLUME 3, STERILIZER
OPERATING PROCEDURES — LABORATORY USAGE