3-7
Service Adjustments and Special Features
Service and Maintenance Procedures
129359-452, Volume 1
3. Set or change utilities control times as follows:
a. Press and release the CHANGE VALUES touch pad
six times
to
advance display to...
DAYS IN WEEK
DAY OF WEEK
b. Enter the number of days in week utilities control is desired (i.e., 5 for
Monday - Friday week; 7 for entire week, etc.).
c. Enter the number of day of the above entered week that it is (i.e.,
Wednesday of a 5 day week would be 3, Friday would be 5, etc.).
d. Press and release the CHANGE VALUES touch pad once more (seventh
time) to advance display to...
SHUTDOWN AT (TIME) A
RESTART AT (TIME) P
e. Move cursor (the blinking display position) to the digit you wish to change
by pressing CURSOR direction pad (left or right).
f. Increase or decrease the value (of blinking digit) by pressing the VALUE
direction pad (up or down).
g. Press SAVE VALUES pad to record the utility control times and to
advance control to READY mode. Printout of cycle values with utility
control times will also be furnished.
3.8
Laboratory Option Description
When the sterilizer is set for laboratory use the operator selects all values
shown in TABLE 3-2 in addition to all the normal hospital values. Also, sterilize
and dry time values are in hours, minutes and seconds instead of only minutes.
Values are set using the same procedure as hospital values. Printouts will
indicate LABORATORY STERILIZER when unit is set for laboratory. (See
printout example.)
Laboratory processes may require additional special sterilizer adjustments
described in Paragraphs 3.6.1, 3.8.2 and 3.8.3.
Refer to VOLUME 3, STERILIZER OPERATING PROCEDURES – LABORA-
TORY USAGE when sterilizer is being used in the laboratory mode of operation.
3.8.1
Set Sterilize Temperature Overdrive
Controls are factory calibrated to control chamber temperature at 3.0 F* (2 C)
above the set temperature. This overdrive setting assures effective load
temperature control and provides for the shortest possible cycle time. For
special sterilizer applications this can be adjusted from 0-9.9 F.