3-4
129359-452, Volume 1
Service and Maintenance Procedures
Service Adjustments and Special Features
• If
correct
access code is entered, the following message is displayed:
CHANGE ACCESS CODE?
NEW CODE = 0000
(Cursor position (first digit) will be blinking on and off.)
NOTE: Maintain a written record of the currently set access code. If access
code is lost or forgotten, have a qualified service technician reset the access
code function.
5. Change access code
if desired
by repeating step 1a through 1h.
. . . or . . .
• Press CHANGE VALUES touch pad
if not changing access code
to
advance control.
NOTE: If a number of cycles are currently locked-out, that number will appear
in the following message instead of “0”. The following message is displayed:
CYCLES TO LOCK OUT = 0
(Cursor position [0] will be blinking on and off.)
NOTE: Locked-out cycles will always be in order; i.e., if one cycle locked-out,
it will be #1; if two cycles locked-out, they will be #1 and #2; etc.
6.
If changing the number of locked-out cycles
, press VALUES “UP”
direction touch pad the number of times equal to the desired number of
locked-out cycles (1, 2, 3 or 4), then press CHANGE VALUES touch pad to
advance control to cycle #1.
. . . or . . .
If not changing the number of locked-out cycles
, press CHANGE
VALUES touch pad to advance control to cycle #1.
7. Change cycles and cycle values per Paragraph 3.3 of VOLUMES 2, or 3 of
this manual.
3.4
Language Option
Sterilizer can be set to print and display in English, French or Spanish.
1. With power switch OFF, set Dip Switch #4 on Display Board to ON position.
Turn power switch ON. Screen should display:
1 ENGLISH
3 ESPANOL
2 FRANCAIS
2. Press appropriate PB to select desired language.
3. Set dip switch to OFF position to retain selected language.
NOTE: If it is desired to operate the unit in multiple languages at various times,
leave the dip switch ON. Each time the unit is powered up, the display will ask
for language choice.
Figure 3-2. Access Code