3-2
129359-452, Volume 1
Service and Maintenance Procedures
Service Adjustments and Special Features
TABLE 3-2 SERVICE ADJUSTMENTS
(Field Test Mode)*
Function
Factory Setting
Optional
Sterilize Temperature
Overdrive*
3.0 F
0-9.9 F
Overtemperature Point
20.0 F
0-99.9 F
Undertemperature Point
2.0 F
0-9.9 F
Pressure Points
P1, P2, P3
26, 1, 3 psig
0-36 psig
Vacuum Point (Express,
Prevac Cycles)
26, 26.5 in Hg
0-36 in Hg
Number of
EXPRESS - 2
0-99
Vacuum Pulses
PREVAC - 4
0-99
Print Intervals
1 min./FLASH
0-99 min.
(Each Cycle)
5 min./others
0-99 min.
* These values are all operator settable in the laboratory mode.
** Sterilize Temperature Overdrive, Overtemperature and Undertemperature are settable
in 1/10 degree increments.
3.1
Automatic Duplicate Print
Sterilizer can be set to automatically furnish a duplicate printout of each cycle
at the end of the cycle. First line will always read . . . DUPLICATE PRINT . . . and
complete printout of cycle data will be furnished.
3.2
Intermittent Buzzer When Cycle Complete
Sterilizer can be set so that the interim buzzer does not sound when cycle is
complete. All additional warning buzzer alarms will still sound even if complete
buzzer is set to OFF.
3.3
Access Code To Set Cycle Values
NOTE: When Access Code function is set to OFF (service adjustment), no
printouts or displays will indicate the feature exists.
1. Open the printer door and position the power switch to ON.
• Display panel lights up.
• The printer records the time and date that the power is turned ON.
NOTE: When power to prevacuum sterilizers is switched on, display will show
the following messages:
•
First Message:
PERFORM LEAK TEST? Press cycle selector touch pad
number 2 for NO.
•
Second Message:
PERFORM DART TEST? (BOWIE-DICK) Press cycle
selector touch pad number 2 for NO.
2. Close the printer door and raise the sliding door over the touch pad panel
to access the VALUES touch pads.
Содержание EAGLE SERIES 3011