3-5
Service Adjustments and Special Features
Service and Maintenance Procedures
129359-452, Volume 1
NOTE: Whether switch is ON or OFF, any time a burn-in occurs, the display will
ask for a language choice.
3.5
Sterilizer Number
Sterilizer can be set to print a specific sterilizer identification number from 00
to 99.
1. While in service test mode (Dip Switch #4 on Display Board), screen
displays:
STERILIZER = VAC00
OR
STERILIZER = GRA00
2. Use arrows to change 00 to desired number. Press SAVE VALUES PB to
save.
3.6
Set Sterilizer Clock/Calendar Time and Date
NOTE: When power to prevacuum sterilizers is switched on, display will show
the following messages:
•
First Message:
PERFORM LEAK TEST? Press cycle selector touch pad
number 2 for NO.
•
Second Message:
PERFORM DART TEST? (BOWIE-DICK) Press cycle
selector touch pad number 2 for NO.
When the sterilizer is set for laboratory use the operator selects all values
shown in TABLE 3-2 in addition to all the normal hospital values. Also, sterilize
and dry time values are in hours, minutes and seconds instead of only minutes.
Values are set using the same procedure as hospital values. Printouts will
indicate LABORATORY STERILIZER when unit is set for laboratory. (See
printout example.)
1. Open the printer door and position the power switch to ON.
• Display panel lights up.
• The printer records the time and date that the power is turned ON.
2. Close the printer door and raise the sliding door over the touch pad panel
to access the VALUES touch pads.
NOTE: If access code feature is enabled, bypass code by pressing CHANGE
VALUES touch pad when NO is blinking.
3. Check display for correct time and date. To change either:
a. Press and release the CHANGE VALUES touch pad until the time and
date are displayed.
b. Move cursor (the blinking display position) to the digit you wish to change
by pressing CURSOR direction pad (left or right).
Содержание EAGLE SERIES 3011
Страница 3: ...Intentionally Blank Page ...