User Guide
Model 4200 Dust & Opacity Monitor
IMPORTANT : Mounting Guidelines
1. The Transmissometer (1) requires approximately 400mm (16in) with
automatic fail safe shutters or 250mm (10in) without automatic fail safe
shutters as a minimum distance beyond the stand pipes for installation (In
addition adequate clearance for access is required for operation).
2. The Retro-Reflector (3) requires approximately 300 (12in) with automatic
fail safe shutters or 165mm (7in) without automatic fail safe shutters as a
minimum distance beyond the stand pipes for installation.
3. Additional clearance is required of approximately 1m (39in) for mounting
and removal procedures.
4. A minimum clearance of 200mm (8in) directly below the instrument is
required for purge air and electrical connections.
5. The maximum recommended air hose length between the blower and the
air purge is 7m (23ft).
Do not exceed this when installing the blower
unit(s).
6. Measurement Pathlength (4) is factory set and cannot be changed by the
user.