Installation
|
3-5
Conductance Limited Enclosed Source and
AMETEK Vacuum Chamber - Figure 3-4
1.
Using a new copper gasket, place the sensor into the AMETEK vacuum
chamber so that the source resides inside the long end of the chamber.
2.
Orient the barrel notch in the 12 o’clock position (Figure 3-3).
3.
Bolt the source flange in place using the (6) ¼-28 x 1” screws and
washer provided. Tighten in a star pattern to ensure a good seal.
4.
Install the appropriate inlet components to the source end of the vac-
uum chamber if they are not already installed. These consist of a 1”
thick VCR flange, a blank flange, a manual high-conductance valve or
electro-pneumatic (EP) high-conductance valve, two copper gaskets, and
fastener hardware. Inlet details
are found in the Chapter 5.
Figure 3-4.
Connecting the
closed/enclosed-
source to AMETEK
vacuum chamber.
Closed Source
Assembly
Detector Flange
Assembly
Barrel Notch
Up
Copper
Gasket
Ion Volume
(closed or enclosed)
AMETEK
Vacuum Manifold
Ceramic
Insulator
Copper Gaskets
VCR Port
(shown capped)
Closed Source
Assembly
Detector Flange
Assembly
Barrel Notch
Up
Copper
Gasket
Ion Volume
(closed or enclosed)
AMETEK
Vacuum Manifold
Ceramic
Insulator
Copper Gaskets
VCR Port
(shown capped)
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...