5-2 | Dycor Dymaxion Mass Spectrometer
Sampling Under Vacuum
If the sample environment total pressure is less than 10
-4
Torr, the mass spectrom-
eter can be directly fitted into the sampling environment without an inlet with a
differential pumping system.
Sampling at Greater than 10
-4
Torr - High Conductance Inlet
If the sampling environment is greater than 10
-4
Torr, an inlet is required. For
these applications, a high conductance inlet (Figures 5-1 and 5-2) can be fitted
to the sample chamber. This inlet consists of a length of tubing installed between
the sample chamber port and the mass spectrometer. Connection of the high
conductance inlet makes it possible for the Dymaxion to sample from the sample
chamber without being directly fitted to the sample chamber.
When the pressure in the sampling chamber is greater than 10
-4
Torr (but be-
low atmospheric pressure), the high conductance valve is closed tightly and the
aperture inlet is opened. The aperture inlet consists of a small aperture within a
valve that connects the sampling system to the vacuum environment of the mass
spectrometer. The aperture inlet should be opened slowly to evacuate the bypass
arm. Once this volume has been evacuated, the pressure in the test chamber, the
diameter of the aperture, and the pumping speed determine the pressure of the
Dymaxion. Aperture diameters and their corresponding sample chamber pres-
sures are provided in Figure 5-3.
Figure 5-1 .
High conductance
inlet with manual
valves.
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...