Service and Parts
|
7-1
BOARD REPLACEMENT
The Dymaxion unit contains five boards (Figure 7-1a) that can be replaced in the
field if damaged or found to be not working properly. These boards can be re-
placed using only a #1 Phillips screwdriver and a 1/4” nut driver. The boards are:
•
Master Board
Located behind the left front panel on the front of the unit.
•
Control Board
Located behind the front heat sink and LED panel on the left side.
•
Scan Board
Located behind the front heat sink and LED panel on the right side.
•
Amp Card
Located behind the rear heat sink and under the Amp cover.
•
RF Board
Located behind the rear heat sink and separated from it by the conductor
guide.
4/40 x 3/16"
screws
4/40 x 1/4"
screws
Small pattern kep nut
must be used
Master Board
Scan Board
Control
Board
Option Plate
Front Heat Sink
Front Panel
Amp Card
RF Card
Scan Card
Control Card
Option Plate
Master
Board
4/40 x 3/16"
screws
4/40 x 1/4"
screws
Small pattern kep nut
must be used
Master Board
Scan Board
Control
Board
Option Plate
Front Heat Sink
Front Panel
Amp Card
RF Card
Scan Card
Control Card
Option Plate
Master
Board
Figure 7-1a.
Dymaxion board
locations.
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...