Maintenance & Troubleshooting
|
6-15
Filament Trip/Overpressure and LEDs
1.
Check the LEDs on the back of the Dymaxion. Filament status can be
determined by the color of the lights. Refer to the previous section for
LED details.
2.
Make sure the filament is turned on by checking the “Light Bulb” icon
on the toolbar in the Dycor System 2000 software to make sure it
is yellow (ON). Check the filament status and emission current by
accessing the Edit menu on the toolbar and selecting “Add Display.”
Click on “Head Status.”
3.
Restart the entire system and recheck the filament status and emis-
sion current.
4.
If there is no filament current, turn off the instrument, pull the Dymax-
ion from the analyzer head and check the continuity between Pins 3
and 8.
5.
If the filament is burned out (LED is red), open the vacuum system and
replace the filament assembly.
If the filament checks out OK, do the following:
•
Check the LEDs on the back of the Dymaxion.
•
If the LEDs are green and there is no overpressure and the filament
has not tripped, call the factory for assistance.
Caution Against Swapping Components
The Dycor System 2000 software has been programmed to accept the combina-
tion of a sensor probe and Dymaxion with different configurations and to auto-
matically set a default configuration for the combined components. However,
when mixing components with different configurations such as a multiplier Dy-
maxion and a Faraday cup quadrupole head, the default configuration is always
the Faraday cup configuration.
• Do not mix components that come from different Dycor models.
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...