3-2 | Dycor Dymaxion Mass Spectrometer
Mechanical Installation
If you are not familiar with high vacuum systems and procedures or
with handling the quadrupole head to prevent contamination, refer
to Chapter 6 of this manual for proper handling procedures before
continuing.
The Dymaxion is shipped wrapped in a protective covering. Check all parts to
ensure that nothing has been damaged. If you discover damage, report it to the
factory immediately.
Visually check the filament for damage.
Install the sensor assembly into vacuum system first.
Connect the Dymaxion electronics to the sensor.
Connect the Dymaxion electronics to 24 VDC power.
Before connecting the 24 VDC power to the electronics, ensure that
the electronics are connected to the sensor. Otherwise, damage to the
electronics is possible.
Always remove 24 VDC power from the Dymaxion electronics before
removing the electronics from the sensor.
If the analyzer is a 230-Volt type, the power cord plug must be
changed to the proper type to match your country and/or electrical
code.
Check Sensor Filament.
1. Before installing the sensor:
Check the filament to ensure that it has not been damaged during shipment.
Visually inspect the filament to make sure that it is not broken.
2. Using a standard Ohm meter, check the continuity of the filament to make
sure that an open circuit does not exist between Pins 3 and 8 of the
vacuum feedthrough pins (see Figure 3-1).
A resistance of < 1 Ohm is required.
3. Check the resistance between Pins 3 or 8 to all other pins. They should all
show an open resistance to operate properly.
NOTE
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...