Installing the new filament
1.
Install the new filament being careful not to bend or break the delicate
filament wires. Handle only the shorting tabs on the filament plate.
2.
Tighten the set screws that hold the new filament in place.
3.
Replace the ion volume by setting it back on the head and secure it
with the set screw.
4.
Remove the shorting tabs on the filament by gently bending them
back and forth with needle nose pliers until the tabs break at the
notches at their bases. Be sure the filament coils are aligned with the
slots in the ion volume and source. If they are not aligned, a minor
adjustment can be made by gently bending the filament plate or by
raising the filament base rods in their barrel connectors.
Be sure the filament coils are aligned with the slots in the ion volume
and ionizer. Make sure the filament coils do not touch the base plate,
ionizer or ion volume in any manner other than at welds.
5.
Install a new copper gasket on the quadrupole feedthrough flange.
Be sure it remains centered on the flange as the head assembly is
installed into the vacuum system using the six bolts.
6.
Tighten the bolts using a cross pattern to ensure proper tightening
and a leak-tight connection. The maximum torque required is 110 in-lb
(9 ft-lb).
7.
Check for electrical isolation between the feedthrough pins.
•
The check should be performed on the pins on the atmosphere side
of the unit (Figure 6-1).
•
The pins should be electrically isolated from each other and from
ground.
•
There should be approximately one ohm continuous circuit between
the filament Pins 3 and 8. They should be isolated from ground.
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...