Overview
|
1-3
ion volume and create ions.
Once the positive ions are formed, they are extracted from the ion region and
focused towards the quadrupole mass filter by a difference in potential. Lens 1
has an applied negative voltage that, due to a difference in electrostatic potential,
attracts the newly formed positive ions, passing them through Lens 2. Lens 2, in
turn, focuses the electrons into the quadrupole.
EI Ionization
The energy of the bombarding electrons is much greater than the bonds that hold
the molecule together. When a single electron is removed, a positive ion is cre-
ated - this is referred to as a molecular ion. When the bombarding electrons cause
the molecular bonds to break, fragment ions form. This ionization process is
known as electron impact ionization (EI).
Operation of the mass spectrometer depends on maintaining low vacuum pres-
sures. These pressures take into account not only the operational pressures of the
mass spectrometer, but also the pressures of the sampling system environments.
Figure 1-1a.
Closed source
components.
Repeller
Filament
Source Grid
Focus Plate
. .
. .
. .
. .
. .
Filament
Repeller
Source Grid
Focus Plate
Top Alumina
Collar
M ass Filter
Figure 1-1b.
Open source components.
Содержание Dycor Dymaxion Mass Spectrometer
Страница 18: ...1 8 Dycor Dymnaxion Mass Spectrometer This page intentionally left blank...
Страница 22: ...2 4 Dycor Dymaxion Mass Spectrometer...
Страница 62: ...5 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 75: ...Maintenance Troubleshooting 6 13 Figure 6 7 Quadrupole head pinout view of flange atmosphere side...
Страница 78: ...6 16 Dycor Dymaxion Mass Spectrometer This page left blank intentionally...
Страница 86: ...7 8 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 90: ...7 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 96: ...A 6 Dycor Dymaxion Mass Spectrometer JP18 JP16 JP17 JP19 JP20 JP21 JP13 JP12 JP14 JP15 JP9 JP6 JP10 JP7...
Страница 102: ...A 12 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...
Страница 117: ...Opto 22 Option Appendix B 15 OPTO 22 MODULE DATA SHEETS AMETEK Part 25540JE Input Module 2 Channel Analog 20 mA to 20 mA...
Страница 118: ...Appendix B 16 Dycor Dymaxion Mass Spectrometer AMETEK Part 25538JE Input Module 2 Channel Analog Type J K E Thermocouple...
Страница 119: ...Opto 22 Option Appendix B 17 AMETEK Part 25539JE Input Module 2 Channel Analog 10 VDC to 10 VDC Voltage Input...
Страница 120: ...Appendix B 18 Dycor Dymaxion Mass Spectrometer AMETEK Part 25541JE Output Module 2 Channel Analog 4 20 mA...
Страница 121: ...Opto 22 Option Appendix B 19 AMETEK Part 25542JE Output Module 2 Channel Analog 0 to 10 VDC...
Страница 122: ...Appendix B 20 Dycor Dymaxion Mass Spectrometer AMETEK Part 25578JE Digital Module 4 Channel AC Input 9 140 VAC...
Страница 123: ...Opto 22 Option Appendix B 21 AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 124: ...Appendix B 22 Dycor Dymaxion Mass Spectrometer AMETEK Part 25577JE Digital Module 4 Channel Isolated AC Output 12 280VAC...
Страница 125: ...Opto 22 Option Appendix B 23 AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 126: ...Appendix B 24 Dycor Dymaxion Mass Spectrometer AMETEK Part 25544JE Output Module 4 Channel Source DC 5 60 VDC...
Страница 127: ...Opto 22 Option Appendix B 25 AMETEK Part 25543JE Input Module 4 Channel 2 5 28 VDC...
Страница 128: ...Appendix B 26 Dycor Dymaxion Mass Spectrometer This page intentionally left blank...