Customer Service 800-528-7411 or 602-470-1414
Page 59 of 68
INSTALLATION of Chlorine or Ammonia Filter Kit:
1. Disconnect the tubing between the 631-X intake and the Vacushield filter from the
Vacushield filter.
2. Connect the Chlorine or Ammonia filter between the Vacushield filter and the intake
of the 631-X as shown in the pictures above and below.
3. Position the filter as shown to determine where to mount the Kwik Clip.
4. Peel the covering off of the back of the Kwik Clip and mount the clip to the cover
below the 651 keypad.
5. Position the Chlorine or Ammonia filter in the Kwik Clip.
6. Use the included Wire Tie to secure the tubing as indicated, but
do not tighten
the
wire tie or the flow will be restricted.
Install Wire Tie around both tubes to secure tubing
Jerome 651 with an Ammonia Filter installed