17 POTENTIAL INTERFERENCES
For some applications, the Jerome Hydrogen Sulfide monitoring system may be
installed in locations where foreign compounds are present. Depending on the foreign
compounds present, in some cases erroneously high readings may occur.
In areas where chlorine gas is present, AMETEK Brookfield recommends the
installation of a Jerome Chlorine Filter Kit (P/N Y990-0200). When installed, this filter
will remove the chlorine gas from the sample stream so that the Jerome monitoring
system can accurately measure levels of Hydrogen Sulfide.
An Ammonia Filter Kit (P/N Y990-0201) is also available for locations where ammonia
may be present.
For pricing and delivery information on the chlorine and ammonia filters, contact Jerome
Sales.
NOTE: PLEASE READ THE DATA SHEETS INCLUDED WITH THE FILTERS FOR
PROPER INSTALLATION AND USE OF THE CHLORINE AND AMMONIA FILTERS.
Filter data sheets are also available online at the website