SYSTEM
DATE/TM
(time)
Edit the current month, day, year, hour, minute, and second.
ID
Settings used to identify the instrument.
•
INSTRMNT - Contains up to 40 characters that are used to
identify the instrument or location.
•
631 SN - Contains the serial number of the installed 631-X
instrument. Do not edit this value unless swapping 631-
X’s.
•
651 SN - Contains the serial number of the instrument. Do
not edit this value.
RADIO
ADDRESS: ##### - The address of the radio modem in this
651. Valid input is 2 to 254. Default is 2.
UP ADDR: The Up Address is set to 1 unless sending packets
thru a repeater before reaching the base station radio; use
repeaters address to send thru a repeater.
RETRIES: The number of times the radio will try to send
information to the base station.
Valid input is 1 to 9. Default is 5.
NET ID: ##### - The radio network ID. All radios on the same
network must have the same ID. Valid input is 0 to 65535.
Default is 1912.
POWER: Set the TX (transmission) power level.
Default is 1W.
RESET
Resets the monitor and clears any errors. Turn off the 631-X
before selecting this option.
TEMP C-F
Select temperature displayed in either
F or
C.
WIND
ENABLE: YES, NO
– Enable wind gauge functionality.
DIR CAL
–
NOTE:
This is not typically used
, but can be used
to get a high degree of accuracy from the wind gauge and
calibrate the directional potentiometer. If done incorrectly, it
could result in significant directional inaccuracy.
The maximum resistance is displayed while the user slowly
rotates the wind vane 2-3 rotations. The maximum
resistance is at 355 degrees. Press [SELECT] to save the
new calibration.
TMP FACT
– Set the compensation value for the external
temperature probe for the effects of the internal box
temperature. Default is 0.11.
Do not edit this value.
FACTORY
(Factory use only
– DO NOT EDIT OR CHANGE DATA OR
SETTINGS)