Section 5 • Infusion History Reports
47
SECTION 5 - INFUSION HISTORY REPORTS
5.1
Pump Infusion History
To obtain the pump infusion history:
Step #1
Place the pump in pause mode.
Step #2
Press and hold the BOLUS button for two seconds until one beep is heard and “HX” is
displayed.
NOTE:
All history will be retained in every mode. When viewing the infusion history,
only the current mode’s history will be viewed, along with the total volume
infused, and the total elapsed time since the last time the pump was cleared; e.g.,
if the pump is in PCA mode, no dose information will be seen. To view specific
history parameters for any mode, the pump must be in that mode.
NOTE:
The remote BOLUS switch may be used to review the infusion history.
5.1.1
PCA Mode Infusion History
INFUSION HISTORY
VISUAL INDICATOR
Current delivery mode
PCA
Total volume infused, including boluses, in ml
“ml” icon
Total number of boluses delivered
No. and
icons
Total number of bolus requests
No. and icons
Elapsed time (time pump has been in run mode since history
was cleared)
and
**
“hr:min” icons
** After 100 hours, pump will only display hours
(not minutes), e.g., 100H